AMPS-1 OPERATING MANUAL
In the past, some patient monitors did not have isolation between their ECG and BP
channels. With these early monitors, ECG, BP and other patient signals shared a
common circuit ground. But when ECG and BP channels were connected to the same
simulator, the signals could appear to fail required tests. However, isolation problems
such as this are generally not an issue with the current generation of patient monitors.
In most cases, the ISOLATION button (located next to the RS-232 port on the bottom
side of AMPS-1) should remain in the
OFF
(out) position. If there seems to be an
interaction when ECG and BP signals of AMPS-1 are simultaneously connected to a
monitor, press the ISOLATION button to the ON (in) position.
Note: since the ISOLATION function reduces battery life, it should
o n ly
b e u s e d w h e n n e c e s s ary
.
Respiration artifact can be selected on all BP channels; RESP ARTIFACT1 has an
amplitude of 5 mmHg, and RESP ARTIFACT2 has an amplitude of 10 mmHg.
6.2 Available Settings
Press ENTER after selecting a parameter to activate that setting.
6.2.1 Transducer Sensitivity
• 5 or 40
ì
V/V/mmHg
6.2.2 Static Levels (all channels)
• -10, -5, 0, 20, 40, 60, 80, 100, 120, 150, 160, 180, 200, 240, 320 and 400 mmHg
6.2.3 Dynamic Simulations (all channels)
• Arterial 120/80
• Arterial 90/40
• Arterial 160/110
• Left Ventrical 120/0
• Right Ventrical 25/0
• Left Atrium 14/4
• Right Atrium 15/10
• Pulmonary Artery 25/10
• Pulmonary Wedge 10/2
• Swan-Ganz Auto
• Swan-Ganz Manual
Invasive Blood Pressure/Chapter 6
#
Page 36
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......