AMPS-1 OPERATING MANUAL
5 Performance Waveforms
5.1 Overview
AMPS-1 offers a number of performance waveforms to evaluate linearity, frequency
response, R-wave detection and other aspects of the performance of patient monitors
and chart recorders.
When Performance Waves are selected, all BP channels are set to a static pressure of
zero mmHg, and the respiration rate is set to zero.
5.2 Available Settings
Press ENTER after selecting a parameter to activate that setting.
5.2.1 Square Wave
• 2 Hz or 0.125 Hz
5.2.2 Triangle Wave
• 2 Hz
5.2.3 Pulse
• 80 msec, 1 Hz (60BPM)
5.2.4 Sine Waves
• 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz.
5.2.5 R-Wave Detector Test
• 60 BPM haver-triangle waveform
• Selectable widths: 8, 12, 20, 40, 60, 80, 100, 120, 140, 160, 180 and 200 msec
Performance Waveforms/Chapter 5
#
Page 31
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......