AMPS-1 OPERATING MANUAL
PACE POS/NEG
=<
PACE PULSE POSITIVE
Figure 51
- PACER POLARITY
Press the UP (
) or DOWN(
) arrow button to scroll through the settings provided
for the pacer pulse width: 0.1, 0.2, 0.5, 1.0 and 2.0 milliseconds.
Pressing the RIGHT (
u
) arrow button will change the menu screen to that of
Figure
50
.
Press the UP (
) or DOWN(
) arrow button to toggle the pacer pulse polarity
between positive-going and negative-going.
Pressing the RIGHT (
u
) arrow button will restore the LCD display to the initial menu
screen of
Figure 47
.
Press the LEFT (
t
) arrow button to scroll the display through the above simulator
parameters in reverse order.
Press ENTER after selecting a parameter to activate your selection.
Pacer/Chapter 8
#
Page 47
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......