AMPS-1 OPERATING MANUAL
10.2.4 Cardiac Outputs
• 3, 4, 4.5, 5, 5.5, 6, 6.5 and 7 liters per minute
10.2.5 Simulation Types
• Trend On/Off
• Slow injection
• Faulty injection
• Left-to-right shunt
• Calibration pulse (1 °C for 1 or 4 seconds)
10.3 Setup
Configure the cardiac output monitor to the following settings:
CATHETER TYPE / SIZE:
Baxter Edwards, 93a 131 7f
CALIBRATION COEFFICIENT:
0.542 for 2 °C Injectate
0.595 for 20 °C Injectate
INJECTATE VOLUME:
10 cc
INJECTATE TEMPERATURE:
2 °C or 20 °C
Connect the Cardiac Output Adapter box (P/N 8000-454) to the AUX input on
AMPS-1, using the 8-pin mini-DIN cable supplied with the adaptor.
Figure 53
- Cardiac Output Adapter
Cardiac Output (Optional)/Chapter 10
#
Page 52
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......