8-2
Valleylab FT10 Energy Platform User’s Guide
General Troubleshooting Guidelines
General Troubleshooting Guidelines
If the VLFT10GEN malfunctions, check for obvious conditions that may have caused the
problem:
•
Check the system for visible signs of physical damage.
•
Make sure the fuse drawer is tightly closed.
•
Verify that all cords are connected and attached properly.
•
If an error code is displayed on the touchscreens, note the code along with all
information on the error screen and use the information in this chapter to proceed.
•
If the problem is still unclear, turn the VLFT10GEN off, wait two minutes before turning
on again.
If the malfunction persists, the system may require service. Contact your institution’s
biomedical engineering department. Covidien Technical Service contact information is
listed on page 9-6.
REM Alarms
If the VLFT10GEN does not sense the correct impedance for the connected REM Polyhesive
patient return electrode, monopolar energy is disabled, the REM indicator on the front
panel illuminates red, the REM symbol appears and enlarges on the touchscreen, and an
alarm tone sounds twice. The REM symbol on the screen returns to its smaller size but
remains red, and RF energy remains disabled until the REM alarm is corrected.
When you correct a REM-alarm condition, the system is enabled and the REM-alarm
indicator on the front panel illuminates green.
Correcting a REM-Alarm Condition
To correct a REM-alarm condition, follow these steps:
1. Inspect the return-electrode plug and cord. If there is evidence of cracks, breaks, or
other visible damage, replace the return electrode and/or the cord.
2. Verify that the patient-return-electrode cord is correctly connected to the energy
platform.
Precaution
Covidien REM Polyhesive patient return electrodes are recommended for use with the VLFT10GEN.
Return electrodes from other manufacturers may not provide proper impedance to work correctly
with the energy platform.
Notice
Alarm characteristics cannot be changed.