10-14
Valleylab FT10 Energy Platform User’s Guide
Standards and IEC Classifications
Guidance and manufacturer’s declaration - electromagnetic immunity
The Valleylab FT10 FT Series Energy Platform is intended for use in the electromagnetic environment
specified below. The customer or the user of the Valleylab FT10 FT Series Energy Platform should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge IEC
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
± 1 kV differential
mode
± 2 kV common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
a
< 40% U
T
(> 60% dip
in U
T
) for 5 cycles
< 70% U
T
(> 30% dip
in U
T
) for 25 cycles
< 5% U
T
(> 95% dip
in U
T
) for 5 sec
0% U
T
for 1 cycle
< 5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
< 40% U
T
(> 60% dip
in U
T
) for 5 cycles
< 70% U
T
(> 30% dip
in U
T
) for 25 cycles
< 5% U
T
(> 95% dip
in U
T
) for 5 sec
0% U
T
for 1 cycle
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Continued