Tr
oubleshooting
Valleylab FT10 Energy Platform User’s Guide
8-9
Correcting Malfunctions
System Alarms
Most system alarms require some action on the user’s part to correct the condition;
however, some are corrected automatically. Use the following list to determine how to
correct an alarm condition.
After correcting the alarm condition, verify that the system completes the self-test as
described in Chapter 3,
.
Error Codes
Error
Code
Error Title
Description
Alarm Type
Clearing Condition
E101
SYSTEM ERROR
FPGA Failure
Safe Mode
Power on reboot,
successful Power-On
Self-Test (POST)
E102
SYSTEM ERROR
CPLD Failure
Safe Mode
Power on reboot,
successful POST
E103
SYSTEM ERROR
FPGA Failure
Safe Mode
Power on reboot,
successful POST
E104
SYSTEM ERROR
Boot Sector
Failure
Safe Mode
Power on reboot,
successful POST
E105
SYSTEM ERROR
RAM Test
Failure
Safe Mode
Power on reboot,
successful POST
E106
SYSTEM ERROR
Processor
Exception
Safe Mode
Power on reboot,
successful POST
E107
SYSTEM ERROR
Unexpected
Interrupt
Safe Mode
Power on reboot,
successful POST
E201
SYSTEM ERROR
Storage Failure
High Alarm
Power on reboot,
successful POST
E202
SYSTEM ERROR
RTOS Failure
High Alarm
Power on reboot,
successful POST
E203
SYSTEM ERROR
OS Failure
High Alarm
Power on reboot,
successful POST
E204
SYSTEM ERROR
Assert Error
High Alarm
Power on reboot,
successful POST
E205
SYSTEM ERROR
Null Pointer
Error
High Alarm
Power on reboot,
successful POST
E206
SYSTEM ERROR
Database Error
High Alarm
Power on reboot,
successful POST
E207
SYSTEM ERROR
Mem Allocate
Error
High Alarm
Power on reboot,
successful POST