Technical Specifications
Valleylab FT10 Energy Platform User’s Guide
10-3
VLFT10GEN Specifications
Environmental Parameters
Input Power
Operation
Transport and Storage
1,2
Ambient
temperature range
50 to 104 °F (10 to 40 °C)
14 to 140 °F (-10 to +60 °C)
Relative humidity
30% to 75% non-condensing
25% to 85% non-condensing
Atmospheric
pressure
700 to 1060 millibars
500 to 1060 millibars
1.
The system can be stored for up to one year without performance degradation upon use.
2.
If the generator is stored at a temperature outside the normal operating range of 10 to 40 °C,
the system is ready for use after at least one hour at ambient temperature 20 °C ± 5 °C.
100–120 VAC
200–240 VAC
Units
Line Ranges
Line Voltage Full Regulation
Range
90–132
180–264
VAC
Line Frequency
47–63
47–63
Hz
Max VA nominal line Voltage
950
950
VA
Max Mains Current
10
6
A
Max Heat Dissipation
180
180
W
Fuses
10
Fuses (2) – 5 mm x
20 mm 10 A, 250 V
fast blow
6
Fuses (2) – 5 mm x
20 mm 6.3 A, 250 V
fast blow
A
Power Cord
NEMA 3-prong hospital-grade connector