Technical Specifications
Valleylab FT10 Energy Platform User’s Guide
10-5
VLFT10GEN Specifications
Internal Memory
Duty Cycle
The VLFT10GEN is capable of operating a duty cycle of 25%, defined as 10 seconds active
and 30 seconds inactive, in any mode for a period of four hours.
Leakage
Storage capacity
8 GB
Precaution
Use of duty cycles greater than 25% (10 seconds active followed by 30 seconds inactive) will
increase the risk that heat build-up under a return electrode may be high enough to injure the
patient. Do not continuously activate for longer than one minute.
Leakage Currents and Patient Auxiliary Currents (IEC 60601-1:2012)
Touch Current
< 100 μA NC, < 500 μA SFC
Earth Leakage Current
< 500 μA NC, < 1000 μA SFC
Patient Auxiliary Current (< 1kHz)
< 10 μA NC, < 50 μA SFC
Patient Auxiliary Current (> 1kHz)
Scaled with frequency per IEC 60601-1:2012, but
does not exceed 10mA NC/SFC
Patient Leakage Current
< 10 μA NC, < 50 μA SFC
Total Patient Leakage Current
< 50 μA NC, < 100 μA SFC
NC—Normal Condition
SFC—Single Fault Condition (as defined in IEC 60601-1:2012)
Total Patient Leakage Current—Measurement of patient leakage current with all patient outputs
connected together
High Frequency Leakage (IEC 60601-2-2)
Bipolar (short leads)
< 68.9 mA
RMS
Monopolar measured directly at the ESU terminals
< 100 mA
RMS
LigaSure/BPR measured directly at the ESU terminals
< 100 mA