Valleylab FT10 Energy Platform User’s Guide
6-5
LigaSure Function Overview
LigaSure Default Settings
LigaSure/Bipolar Receptacle
The LigaSure/Bipolar receptacle is located adjacent to the lower-right quadrant of the
screen. This receptacle accepts LigaSure instruments and can read bar codes on the RFID
chip or the LigaSure connector.
Output or Instrument
Default
LigaSure
Handswitching—enabled (On)
LF5544 Pistol Grip
•
MONOPOLAR mode—VALLEYLAB
–
15 W
LF5637 and LF5644
•
MONOPOLAR mode—VALLEYLAB
•
Monopolar power
–
CUT—15 W
–
VALLEYLAB—15 W
–
COAG—15 W
Warning
Electric Shock Hazard
•
Do not connect wet instruments to the energy platform.
•
Ensure that all instruments are correctly connected and that no metal is exposed at any
connection point.
Connect instruments to the proper receptacle. Improper connection may result in inadvertent
instrument activation or other potentially hazardous conditions. Follow the instructions provided
with LigaSure instruments for proper connection and use.
Each instrument receptacle on the energy platform is designed to accept only one instrument at a
time. Follow the instructions provided with electrosurgical instruments for proper connection and
use.
Precaution
Read the instructions, warnings, and precautions provided with LigaSure instruments before use.
Specific instructions are not included in this manual.
Inspect instruments and cords—especially for laparoscopic/endoscopic instruments—for breaks,
cracks, nicks, and other damage before every use. If damaged, do not use. Damaged instruments
or cords may result in injury or electrical shock to the patient or surgical team.