Valleylab FT10 Energy Platform User’s Guide
5-3
Rear Panel Bipolar Features
Bipolar
Rear Panel Bipolar Features
Bipolar foot-pedal receptacle
ཱ
Power cord receptacle
Bipolar Quick Setup Instructions
If you are familiar with the VLFT10GEN, follow this abbreviated procedure to set up the
system for bipolar surgery.
If you are not familiar with the VLFT10GEN, refer to the following sections in this chapter
for detailed instructions.
1. Plug the system power cord into the rear panel receptacle.
2. Plug the system power cord into a grounded wall receptacle.
3. Turn on the system and verify that the power-on self-test successfully completes.
4. If using a foot pedal, connect it to the Bipolar foot-pedal receptacle on the rear panel.
5. Connect the instrument to the Bipolar instrument receptacle on the front panel.
6. Verify or change the mode and power settings.
Important
Do not plug the system power cord into a power strip or extension cord.
Warning
Use only Covidien foot pedals. Use of other manufacturer’s foot pedals is not recommended;
unexpected output may occur.
Warning: Risk of Fire.
Replace Fuse as Marked
250V, F10.0A (100-127Vac)
250V, F6.3A (220-240Vac)
Avertissement: Risque du feu.
Remplacez les fusibles
comme marqués.
250V, F10.0A (100-127Vac)
250V, F6.3A (220-240Vac)
Monopolar 2
Monopolar 1
Bipolar
2
Monopolar
1
Monopolar
1
2