Technical Specifications
Valleylab FT10 Energy Platform User’s Guide
10-13
Standards and IEC Classifications
The essential performance requirement per IEC60601-1 does not apply to VLFT10GEN.
Basic safety is the performance requirement used during immunity testing.
The system intentionally applies RF energy for diagnosis or treatment during activation. Observe
other electronic medical equipment in the vicinity during the system activation for any possible
adverse electromagnetic effects. Ensure adequate separation of electronic medical equipment
based on observed reactions.
The use of accessories, other than specified in the
Valleylab FT10 Energy Platform User Guide and
Service Manual
, may result in increased emissions or decreased immunity of the system.
Guidance and manufacturer’s declaration - electromagnetic emissions
The Valleylab FT10 FT Series Energy Platform is intended for use in the electromagnetic
environment specified below. The customer or the user of the Valleylab FT10 FT Series Energy
Platform should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic
environment - guidance
RF emissions
CISPR 11
Group 2
The Valleylab FT10 FT Series
Energy Platform must emit
electromagnetic energy in
order to perform its intended
function. Nearby electronic
equipment may be affected.
RF emissions
CISPR 11
Class A
The Valleylab FT10 FT Series
Energy Platform is suitable for
use in all establishments other
than domestic and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Notice