LigaSur
e
Valleylab FT10 Energy Platform User’s Guide
6-9
LigaSure Instruments
Hand-Activation ON/OFF Switch
Some LigaSure instruments cause the hand-activation icon to appear in the lower left
corner of the LigaSure control panel. The hand-activation switch toggles between the ON
and OFF setting.
For example, the LF1212 Curved, Small Jaw, Open Sealer/Divider, is a LigaSure instrument
that can be used with hand-activation on or off. When hand-activation is off, a foot pedal
is used to initiate energy delivery.
Hand activation enabled
In this example, the enabled hand activation appears in the lower-left corner of the
LigaSure quadrant. Touching the
ON/OFF
switch turns off the hand activation, requiring
the use of a foot pedal.
Invalid Instrument
If the VLFT10GEN does not recognize the attached instrument, a pop-up will display the
words “I
NVALID
I
NSTRUMENT
.” Refer to the following steps to resolve the issue.
1. Confirm that a LigaSure instrument is in use.
2. Reconnect the instrument using firm pressure to insert the instrument into the
LigaSure/Bipolar instrument receptacle.
3. If “I
NVALID
I
NSTRUMENT
” continues to appear in the status bar, use a new LigaSure
instrument or electrode.
1