![Covidien Valleylab FT Series Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/covidien/valleylab-ft-series/valleylab-ft-series_user-manual_2673869136.webp)
10-6
Valleylab FT10 Energy Platform User’s Guide
VLFT10GEN Specifications
Radio Frequency Identification (RFID)
The RFID module is located above the LigaSure/Bipolar receptacle. The intended use of the
RFID module is to identify the inserted LigaSure instrument and configure the generator
with the data included in the RFID tag.
Transmit/Receive Frequency Range
13.56MHz
RF Output Power
68.17 dBu V/m @3 meter
Type of Antenna
Integral Loop Antenna
Modulation
Modulation: Amplitude-shift keying (ASK)
Mode of Operation (Simplex / Duplex)
Duplex
Contains Module FCC ID
2AAVI-JDK1901
Contains IC ID
11355A-JDK1901