6-8
Valleylab FT10 Energy Platform User’s Guide
LigaSure Instruments
Connecting Dual-Plug LigaSure Instruments
LigaSure connector (purple)
ཱ
LigaSure/Bipolar instrument receptacle
ི
Monopolar connector (blue)
ཱི
Monopolar 2 instrument receptacle
ུ
LigaSure touchscreen quadrant
ཱུ
Monopolar 2 touchscreen quadrant
ྲྀ
REM patient return electrode
Setup
1. With the bar code on the purple LigaSure connector
(1)
facing up, firmly insert it into
the LigaSure/Bipolar receptacle
(2)
next to the LigaSure touchscreen quadrant
(5)
.
2. With the bar code on the blue Monopolar connector
(3)
facing up, firmly insert it into
the Monopolar 2 receptacle
(4)
next to the Monopolar 2 quadrant
(6)
.
The energy platform reads the information from the two connectors and sets the
power output accordingly:
•
LigaSure
– Power output is selected automatically and cannot be changed from
the LigaSure touchscreen.
•
Monopolar
– Power output is set to the default values and can be edited in the
Monopolar 2 quadrant.
.
1
2
3
4
5
6
7