Technical Specifications
Valleylab FT10 Energy Platform User’s Guide
10-11
Standards and IEC Classifications
Standards and IEC Classifications
The VLFT10GEN meets all pertinent clauses of the IEC 60601-1 editions 2.0 and 3.1, IEC
60601-2-2 editions 4.0 and 5.0, IEC 60601-1-2 editions 2.1 and 3.0 and 4.0, and IEC
60601-1-8 edition 2.1.
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation failure
because of the way in which they are connected to the protective earth conductor.
Type CF Equipment/Defibrillator Proof
(IEC 60601-1, IEC 60601-2-2, and ANSI/AAMI HF18)
This VLFT10GEN provides a high degree of protection against electric shock,
particularly regarding allowable leakage currents. It is type CF isolated
(floating) output and may be used for procedures involving the heart.
Foot pedal
Volume adjustment for activation tones
Equipotential grounding point
Equipment should not be disposed in trash
Do not reuse (Single use only)