Monopolar
Valleylab FT10 Energy Platform User’s Guide
4-13
Monopolar Electrodes
When an instrument with VALLEYLAB-mode capability is attached to the Monopolar 2
receptacle, the energy platform detects the instrument type and changes the screen from
the default two-button to the corresponding screen.
Force TriVerse Screen
Power bar indicator
ཱ
Slider position indicator
ི
CUT mode ON/OFF switch
ཱི
CUT mode power indicator (watts)
ུ
VALLEYLAB mode power indicator (watts)
ཱུ
COAG mode power indicator (watts)
Power Bars—
The five power bars represent the range of power available for the
instrument attached to the energy platform receptacle. The system automatically identifies
the instrument and selects the default power-bar setting. The power bars represent power
available
to the instrument and can be set independently from the slider bars indicated in
the slider-position indicator. Power bars can only be changed at the touchscreen interface
on the energy platform.
Note:
Refer to the individual instrument instructions for power-bar output in watts.
ON/OFF Switch for CUT—
Enable or disable CUT. CUT will not be available from the
yellow button on the instrument when disabled on the touchscreen.
Slider Position (if available on the instrument)—
The energy platform detects the
position of the instrument’s slider switch, and the slider position indicator displays this
position. Slider position can only be changed by the instrument user in the sterile field.