Valleylab FT10 Energy Platform User’s Guide
General Warnings and Precautions
Warnings for Implanted Electronic Devices (IEDs)
IEDs include, but are not limited to, pacemakers, neurostimulators, implantable
cardioverter defibrillators (ICDs), ventricular assist devices (VAD), spinal cord stimulators,
cochlear implants, infusion pumps, and bone growth stimulators.
Post Surgery Safety Issues
Use only instruments that can withstand the maximum output (peak) voltage for each output
mode as listed in Chapter 10,
. Using an instrument with a voltage rating
that is lower than the maximum output voltage may result in injury to the patient or the operator,
or damage to the instrument.
Information on voltage ratings for non-Covidien instruments should be obtained from the
instrument’s manufacturer.
Notice
All Covidien instruments have voltage ratings that are greater than the maximum output voltages
in the VLFT10GEN and are thus fully compatible.
Inspect instrument plugs for wear before every use. Worn plugs may result in a loose or stuck
connection to the generator.
Warning
Use the system with caution in the presence of internal or external pacemakers or other implanted
devices. Interference produced by electrosurgical equipment can cause a pacemaker or other
device to enter an unsafe mode or permanently damage the device. Consult the device
manufacturer or responsible hospital department for further information when use is planned in
patients with implanted medical devices.
Warning
Shock Hazard
Before cleaning or servicing the unit, disconnect the power plug from the power
outlet in order to completely isolate the generator from mains power.
Notice
Do not clean the energy platform with abrasive cleaning or disinfectant compounds, solvents, or
other materials that could scratch the panels or damage the energy platform.