Valleylab FT10 Energy Platform User’s Guide
6-11
Activating the LigaSure Instrument
Incomplete Seal Cycle - Short
The generator detected a short circuit between the LigaSure jaws—possibly due to contact
with a metal object or excess fluid around the surgical site.
If this message appears the user should
1. Release the foot pedal or handswitch button.
2. Check for metal clips or staples.
3. Reposition the instrument to avoid contact with metal.
4. Remove excess fluids from the surgical site.
5. Reactivate the seal cycle.
Incomplete Seal Cycle - Open Circuit
The LigaSure jaws were not fully closed or the grasped tissue was too thin.
If this message appears, the user should
1. Release the foot pedal or activation button.
2. Open the instrument jaws and inspect for a successful seal.
3. Clean and dry the LigaSure jaws.
4. If possible, regrasp thicker tissue.
5. Fully close the instrument jaws.
6. Reactivate the seal cycle.