CI Systems
SR-800R
8.6 Hood
8.6.1 Overview
An optional Hood can be supplied separately. When mounted on the Blackbody and
the UUT is placed in front of the Hood (with minimal air gap), dry Nitrogen is
pumped to create a dry atmosphere with minimal turbulence (minimal pressure to
create a positive pressure atmosphere).
The Hood main components are:
Dry Nitrogen Inlet
(1 bar max): the port to which the Dry Nitrogen is connected
when using a Hood (instead of using the Blackbody Dry Nitrogen Inlet).
Latch
: four latches, used to attach the Hood to the Blackbody.
Dry Nitrogen Exhaust Slots
: nipples, located around the Hood inner-side. The
nipples are used to flow the Dry Nitrogen around the Blackbody.
Figure 8-11: The LT Blackbodies Hood - Main Components
Dry Nitrogen Inlet
(1 bar max)
Latch (x4)
Inner-side View
Dry Nitrogen
Exhaust Slots
605-7040
Operation
Manual
Rev D
8-19
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com