CI Systems
SR-800R
3. Remove the sensor and store it in its package to avoid damage.
4. Clean the Sensor port using the plastic straw that has been supplied:
a. Insert the plastic straw into the Sensor port.
b. Rotate the plastic straw several times in the Sensor port.
c. Remove the plastic straw from the Sensor port.
7.3.1.2 Installation
1. Apply a thin layer of thermal grease (supplied with the kit) on the bottom section
of the sensor (approx. 1cm on the free end of the sensor), using a toothpick (see
Figure 7-6
).
2. Insert the sensor into the Emitter port until the magnet secures it.
3. Connect the Sensor male connector to the Blackbody (see
Figure 7-9
).
605-7040
Operation
Manual
Rev D
7-11
June 2013