CI Systems
SR-800R
6.2.25 ISTEMPERATURESTABLE
Summary
: Use this command to monitor whether the temperature has reached a
stable condition.
Note
: If this command is called immediately after calling the SetTemperature or
ST commands, then it may not show that the temperature is not stable. Wait one
(1) second before using this command.
I S T E M P E R A T U R E S T A B L E
E
O
M
Input Operand
: None.
Return Values
:
•
0 = No
•
1 = Yes
6.2.26 RD
Summary
: This command is used to get the current differential temperature, i.e.,
the difference between the emitter and the ambient (Target) temperature.
R D
E
O
M
SR-80 command format compliance.
Input Operand
: None.
Return Values
: Current Differential Temperature with 3-digit accuracy.
6.2.27 RE
Summary
: This command is used to get the current emitter temperature.
R E
E
O
M
SR-80 command format compliance.
Input Operand
: None.
Return Values
: Current Emitter Temperature with 3-digit accuracy.
605-7040
Operation
Manual
Rev D
6-23
June 2013