CI Systems
SR-800R
7.4
Updating the Calibration Tables
This section provides step-by-step instructions on how to download the new replaced
sensor calibration data into the blackbody.
7.4.1 Required
Equipment
A PC with WIN-XP Operating System.
One of the following communication methods: an available RS232 port, Ethernet
port or NI GPIB XP card installed in the PC.
7.4.2 Prerequisites
Install the Removable Sensor Utilities program:
1. Launch the Removable Sensor Utilities installation CD.
2. Follow the prompts (do not change all default settings). At the end of installation,
the Removable Sensor Utilities icon is added to your Desktop.
7.4.3 Procedure
1. Preparations:
a. Pending on your preferable communication protocol (RS232, Ethernet or
GPIB), connect the SR-800R Controller to the PC.
b. Switch ON the SR-800R Controller (see
Figure 3-1
) and wait until the
completion of the BIT sequence.
c. Switch ON the PC.
d. Launch the Calibration CD (the Calibration CD is found in the supplied
Removable Sensor Kit).
605-7040
Operation
Manual
Rev D
7-14
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...