6.2.21 *IDN?
Summary
: This command is used to display the Controller Identification from a
GPIB communications device. A controller will respond to this command by
sending the following information: Controller Model (SR-800R, SR200, etc.),
Controller Serial Number, and its Software Version.
Note
: This command is a
GPIB
specific command which when used, interrogates
the GPIB Bus to identify controllers that are connected in the system. This will not
work for RS232/Ethernet communications mode.
Reference Command
: None
* I D N ?
E
O
M
Input Operand
: None
Return Values
: Returns Controller Identification with its Serial Number.
6.2.22 GETTARGETTEMPERATURE
Summary
: This command is used to return the current Target Sensor
Temperature. When no target exists in the system then an
Invalid Operand
Message is returned.
This command replaces the RT command.
G E T T A R G E T T E M P E R A T U R E
E
O
M
Input Operand
: None.
Return Values
: Target Sensor Temperature with 3-digit accuracy.