CI Systems
SR-800R
8.4.4 Cable-Connection
1. Connect the Head cable between the Blackbody and the SR-800R Controller (see
Figure 8-1
).
2. Connect the Julabo Circulation System cables (see
Figure 8-9
):
a. Connect
the
RS-232 Cable
between the SR-800R Controller and the Julabo
SERIAL
cable-connector.
Note that the Circulator provides three possibilities for the setpoint selection,
i.e., Temperature Setting.
At CI-Systems, the Circulator is set to work in the RS-232 mode with the
following communication setup (for reference only):
•
Baud - 9600
•
Parity - None
•
Stop Bit - 1
•
Data Length - 8 Bit
•
Handshake - None
NOTE
Depending on the SR-800R Controller temperature setpoint, the
only way to control the Julabo Circulation System temperature
is by the SR-800R Controller and via the RS-232
communication port. Therefore, the SR-800R Controller RS-
232 port cannot be used for other purposes.
b. Connect
the
Julabo Control Cable
between the Julabo Controller and the
ALARM
cable-connector.
c. Connect the Julabo Controller Power Cable to the
Circulator Power cable-
connector
.
d. Connect
the
Main Power Cable
to the Mains (either directly or via a 100
VAC transformer).
605-7040
Operation
Manual
Rev D
8-13
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com