3. Enter
the
Slow
and
Fast
arrow Increments (any decimal value).
4. Enter
the
Stabilization Window
values.
NOTES
- The
Slow
Increments
value must be lower than the
Fast
value.
- It is recommended that the
Stabilization
Window
value should
be higher than the
Display Resolution
value.
- The
Slow Increment
value must be higher than the
Display
Resolution
value.
- The
dT MRTD
field is used for MRTD stabilization only.
5. To restore the temperature window values press
(the default values are
as shown in
Figure 3-47
).
6. Press
to update the settings in the system.
7. To close and return to the last opened window (the
Options
window) press
.
8. To close and return to the Main window press
.
605-7040
Operation
Manual
Rev D
3-46
June 2013