CI Systems
SR-800R
6.2.7 GETACTIVEAXIS
Summary
: This command retrieves the current active axis String ID in the
system. There can be only 1 active axis in the system at any given time.
Note
: Use the
SetActiveAxis
command to select the active axis.
Reference Command
: SetActiveAxis
G E T A C T I V E A X I S
E
O
M
Input Operand
: None
Return Values
: ActiveAxisID - return the Active Axis String ID. This can be one
of the following strings; 'Axis1', 'Axis2', 'Axis3', or 'Axis4'.
If there are no valid
axes in the system, then it will return an operand error
.
6.2.8 GETNUMBEROFAXES
Summary
: This command retrieves the number of axes in the system and their
string IDs. It can send up to 5 comma-delimited parameters. The first parameter is
the number of axes in the system and the rest are the String IDs for each valid
axis. It will always respond with at least one parameter.
If there are no axes in the system, then the command returns 0.
Reference Command
: GetActiveAxis, SetActiveAxis
G E T N U M B R O F A X E S
E
O
M
Input Operand
: None
Return Values
:
•
Return1
- Number of axes in the system. For none, return 0. There can be up to
4 axes in a system.
•
Return2-5
-
ActiveAxisID
s for each axis in the system.
Example
: assume system with two axes and they are connected to Axis1 and
Axis2; (Target Wheel and Filter Wheel), then the command will respond:
2
,
Axis1
,
Axis2
605-7040
Operation
Manual
Rev D
6-13
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...