To Change/Update the Size of an Existing Target
:
a. Select a target from the list (you may browse the list by using the
Up / Down
buttons).
b. Press
the
Size
field
. A floating Keypad opens.
c. Using the Keypad enter the new target size (mm), and press
. The
floating Keypad closes, and the new target size is displayed in the
Size
field.
d. Press
, and confirm the modified Target Setup.
5.
To Change/Update the Position of an Existing Target
:
a. Select a target from the list (you may browse the list by using the
Up / Down
buttons).
b. Press
the
Position
field
. A floating Keypad opens.
c. Using the Keypad enter the new target position (MSteps), and press
.
The floating Keypad closes, and the new target position is displayed in the
Position
field.
d. Press
, and confirm the modified Target Position.
e. To verify the target new positioning press
and check that the target
wheel rotates to the desired location.
6.
To Set the Temperature Offset
: The temperature offset settings should be
carried out for every target. If the
Temperature Offset = 0
, then the system is
using the
Radiometric Offset
value, providing that the Radiometric Offset is set
to
On
(see Section
3.2.3.1
, step
4
):
a. In
the
Differential Mode
window set the
DT
to
0
and select a target (see
Section
3.2.2
).
b. Aim a FLIR towards the Focal Plane/Blackbody and check for differences. If
there are any, change the DT until no differences are observed.
c. Record
the
DT
value.
d. Open
the
Target Setup
window (see
Figure 3-35
).
e. Select the same target as selected in step
a
and enter the recorded
DT
value in
the
Temperature Offset
field (recorded in step
c
).
f. Click
.
g. Repeat
steps
a
-
f
for the entire targets.
605-7040
Operation
Manual
Rev D
3-32
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...