6.1.3
Transferred Data Format
RS232: Each byte will be transferred serially as follows:
•
1 START bit.
•
8 data bits (transferred data), LSB (D0) first.
•
1 STOP bit.
•
No parity.
Ethernet: Data is sent in packets. The protocol implemented is based on TC/PIP.
The SR-800R uses a built-in server port of 5200 to respond to client requests. For
customers interested in using Ethernet, CI can provide a sample program
demonstrating the operation of the protocol.
GPIB: DATA is sent in a Byte-Wide bus via IEEE-488 Control Connections.
6.1.4 Naming
Bits
The Most Significant Bit (MSB) in each byte is named D7.
6.1.5
Command Message Structure
Each Command contains a
CommandBody
and a
CommandOperandBody
:
The
CommandBody
and
CommandOperandBody
are separated by a
Space
ASCII Text Character, 0x20 Hex, symbolized by
Space
sign
.
The
CommandOperandBody
contains the command operands. There may be
from no operands to a few operands in the CommandOperandBody. Each Operand
is separated by a
Comma
ASCII Text Character, 0x2C Hex. This is symbolized
by the
Comma
sign.
Commands are case insensitive unless otherwise noted.
605-7040
Operation
Manual
Rev D
6-2
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...