CI Systems
SR-800R
Message
No.
Failure Message
Probable Cause of Failure/
Action
33
"WARNING: The Control Master cannot reach
the required differential temperature (dT),
because the absolute Te temperature value cannot
be below XX
1
degrees. You must select a higher
dT value."
Re-enter new dT values.
34
"ERROR: The Black Body XX
1
temperature has
passed the absolute maximum value of XX
1
. The
Power Amplifier has been shut down. Turn off
the system and wait at least 5 minutes before
powering up again."
Follow the on-screen text and
instructions.
49
"ERROR: Motor is not found!!."
Verify the AXIS cable
connections.
The motor is malfunctioned.
50
"ERROR: Motion Control Card Detection
Failure. All the axes are disabled. Call CI-
Systems for assistance."
Contact CI Systems technical
support.
51
"ERROR: VCO frequency measurement failed.
All the axes are disabled. Call CI-Systems for
assistance."
Contact CI Systems technical
support.
52
"ERROR: VCO frequency is out of range. All the
axes are disabled. Call CI-Systems for
assistance."
Contact CI Systems technical
support.
54
"ERROR: MCC Card axis XX
1
test failed. This
axis is disabled."
Contact CI Systems technical
support.
57
"ERROR - Could not establish communications
with the Head. You must power down the unit
and check the cable to the Head. Contact CI for
further assistance if this error continues to
appear."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
58
"ERROR: The Black Body Head Model Type is
not defined properly. Contact a CI Qualified
Technician."
Contact CI Systems technical
support.
605-7040
Operation
Manual
Rev D
5-3
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...