CI Systems
SR-800R
Message
No.
Failure Message
Probable Cause of Failure/
Action
163
"WARNING: The HEAD SELF CALIBRATION
is Out of Tolerance. Reboot the controller.
If the warning re-occurs, contact CI for further
assistance."
The ambient temperature
changing too fast.
Contact CI Systems technical
support.
164
"WARNING: PID Coefficient Table does not
cover the full operating range of the currently
defined Head.
Contact CI for further assistance."
Contact CI Systems technical
support.
167
"ERROR: Slow increment is lower than display
resolution, Please fix this value"
Enter slow increment value
that is higher than the display
resolution.
169
"WARNING: Stabilization window is lower than
the display resolution. Please fix this value"
Enter stabilization window
value that is higher than the
display resolution.
605-7040
Operation
Manual
Rev D
5-6
June 2013