No.
Failure Message
Probable Cause of Failure/
Action
82
"ERROR: Power amplifier Over Temperature!!
The Power amplifier card temperature has passed
XX
1
[
˚
C]. The temperature control is disabled."
Possible causes are:
1. The vent holes of the Control Master enclosure
are blocked.
2. The Controller fan is not rotating properly. Fix
the problem and restart the system."
Contact CI Systems technical
support.
95
"ERROR: Controller fan malfunction. The
temperature control is disabled. Please fix the
problem and restart the system."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
96
"ERROR: The Power Amplifier VCChigh
Voltage is too high (VCC=48.4V,
VCCmax=47.0V).
The temperature control is disabled."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
97
"WARNING: The Power Amplifier VCClow
Voltage is out of range (VCC=25.6V,
VCCmin=17.0V, VCCmax=23.0V).
The system may not work properly."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
98
"WARNING: Invalid Power Amplifier output
voltages (V1=16.1V, V2=29.1V, VCC=37.3V,
Dvmin=2.0V, Dvmax=8.5V).
The system may not work properly."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
107
"WARNING: Check the cable connection of the
Axis."
Verify the AXIS cable
connection.
117
"WARNING: You cannot run a MRTD Test. The
Axis is not a Target Wheel."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
148
"ERROR: one or more Axis3 parameters are
incorrectly defined. The axis will be disabled."
Follow the on-screen text and
instructions.
Contact CI Systems technical
support.
605-7040
Operation
Manual
Rev D
5-5
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 129: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...