CI Systems
SR-800R
5.5 Fuse
Replacement
WARNING
The fuse is powered even when the SR-800R is OFF. Before
replacing the fuse, power-off the SR-800R and disconnect the
power cable from mains. Failure to comply may result in
injury to personnel and/or damage to equipment.
Tools
:
Flathead Screwdriver.
Parts
:
Fuse, type: F12.5A, size: 20/5mm.
Procedure
(see
Figure 5-1
):
1. Power-off the SR-800R
.
2. Disconnect the power cable from mains.
3. Remove the fuse housing using a Flathead screwdriver.
4. Replace the fuse.
5. Re-install the fuse housing and tighten using a Flathead screwdriver.
6. Reconnect the power cord to mains.
7. Power-on the SR-800R.
Figure 5-1: Fuse Replacement
110/230V
Fuse Housing
(F12.5A, 20/5mm)
605-7040
Operation
Manual
Rev D
5-8
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание ControlMaster SR-800R
Страница 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...