Chapter XVIOscilloscope Probe (Option)
215
Power adapter
Hook
BNC output
cable
Power indicator
Attenuation
switch
Probe body
Power switch
Adapter
access
Hook access
cable
Fig. 16.29 Diagram of P7100 Probe Accessories
Warning:
The maximum input voltage of the P7100 high voltage differential probe is 7kV
(DC+ACpp). Do not exceed this voltage in use. Otherwise, it will cause personal
hazard or injury or damage to the DUTs.
16.12.3 Main Parameters of Probe
The main parameters of the P7100 probe are shown in the table below.
Table 16.12 P7100 Probe Parameters
Model
P7100
Bandwidth (-3dB)
100MHz
Precision
± 1%
Attenuation ratio
1/100, 1/1000
Termination
Between input end and ground: 10MΩ
Input capacitance
Between input end and ground: 2pF
Maximum differential measurement
voltage
1/100: 0.7kV (DC+ACpp)
1/1000: 7kV (DC+ACpp)
CMRR
60Hz:
≥
80dB
1MHz:
≥
50dB
Output voltage (50kΩ load)
± 7V
Power adapter output voltage
DC6V, 700mA
Power adapter input voltage
AC100 ~ 240V, 50 ~ 60Hz
Содержание 4456 Series
Страница 1: ...4456 Series Digital Phosphor Oscilloscope User Manual China Electronics Technology Instruments Co Ltd...
Страница 10: ...VIII...
Страница 55: ...Chapter V Trigger System 45 Fig 5 51 RF Test Signal Example...
Страница 69: ...Chapter VII Mathematical Operation 59 Fig 7 10 Advanced Math Example...
Страница 71: ...Chapter VIII Display System 61 Fig 8 3 Wfm Palette Menu Fig 8 4 Normal Palette...
Страница 72: ...Chapter VIII Display System 62 Fig 8 5 Inverted Palette Fig 8 6 Temperature Palette...
Страница 75: ...Chapter VIII Display System 65 Fig 8 12 XY Display Example...
Страница 165: ...Chapter XI Protocol Analyzer Option 155 Fig 11 242 1553 Trigger and Analysis Example...