Chapter VII Mathematical Operation
56
Fig. 7.5 Configure FFT Menu
Vertical Scale: After clicking the menu key, rotate the Multipurpose b to set the vertical scale of the FFT
waveform.
Vertical Offset: After clicking the menu key, rotate the Multipurpose b to set the vertical offset of the FFT
waveform. Attention: The FFT waveform zero position is at the top of the screen waveform area.
Vertical Units: Select the FFT operation unit, including V, dBV, dBmV, mV.
Center Freq: Set the corresponding frequency of the center of the screen, which can be input by Multipurpose b or
digital keyboard.
Freq Width: Set the frequency width of a screen, which can be input by Multipurpose b or digital keyboard. In this
case, the frequency of the left-most position of the screen is: Center Freq - Freq Width/2, and the frequency of the
right-most position of the screen is: Center Freq + Freq Width/2.
Fig. 7.6 FFT Example