Chapter XVIOscilloscope Probe (Option)
195
Probe front end:
Fully enclosed
design for safer
use
Compensation
adjustment:
For frequency
compensation
in X10 state
Hook: Use a retractable
probe to make your
measurements easier and
safer
Ground clamp:
Use a crocodile-
type ground
clamp to reliably
ground the probe
Adjustment rod:
Adjust the
compensation of
the probe
Color ring sleeve: Put the color ring sleeve on
both ends of the probe to easily distinguish the
probes used in different channels
Automatic identification end:
Automatically identify the probe and
set the impedance of the oscilloscope
and the probe attenuation
Grounding spring:
Shortest grounding
spring for high
frequency testing
Spare needle: Spare
needle, which can be
replaced by users
Needle protective
cap: Needle front
end report
Fig. 16.11 Diagram of P9600A Probe
Warning:
The maximum input voltage of the P9600A passive voltage probe is 300V
(DC+ACpk). Do not exceed this voltage in use. Otherwise, it will cause personal
hazard or injury or damage to the DUTs.
16.4.2 Main Parameters of Probe
The main parameters of the P9600A probe are shown in the table below.
Table 16.4 P9600A Probe Parameters
Model
P9600A
Bandwidth
600MHz
Termination
10M
Ω
±2%
Input capacitance
12pF±2pF
Attenuation ratio
10:1
Maximum input voltage
ACpk
Capacitance compensation range
6-22pF
Probe length
120±2cm
Safety
Complies with EN61010-031 CATII
Содержание 4456 Series
Страница 1: ...4456 Series Digital Phosphor Oscilloscope User Manual China Electronics Technology Instruments Co Ltd...
Страница 10: ...VIII...
Страница 55: ...Chapter V Trigger System 45 Fig 5 51 RF Test Signal Example...
Страница 69: ...Chapter VII Mathematical Operation 59 Fig 7 10 Advanced Math Example...
Страница 71: ...Chapter VIII Display System 61 Fig 8 3 Wfm Palette Menu Fig 8 4 Normal Palette...
Страница 72: ...Chapter VIII Display System 62 Fig 8 5 Inverted Palette Fig 8 6 Temperature Palette...
Страница 75: ...Chapter VIII Display System 65 Fig 8 12 XY Display Example...
Страница 165: ...Chapter XI Protocol Analyzer Option 155 Fig 11 242 1553 Trigger and Analysis Example...