59
Note: The method test screen appears indicating
the power, temperature, volume and time.
Note: Press the operation key below VOLUME
to edit or enter a sample volume.
15.
If Autotrack is turned off, a sample identifi-
cation name or number is required. Press
the operation key below ID.
16.
Using the operation keys below the arrows,
position the cursor on (select) the first
number or letter of the sample identifi-
cation. Press ENTER. Continue to position
the cursor on each letter or number and
press ENTER until the identification is
complete.
Note: If using an identification number only, use
the numeric keys to enter the number
17.
Press READY to return to the test screen.
18.
Lift the instrument cover. Place the filter
on the balance pan. Close the instrument
cover.
19.
Press TARE.
20.
Wait for the instrument to tare the weight
of the filter.
XXXXXX
Power:
xxx %
Temp (xxx): xx.x
Time:
xx:xx
Volume:
xxx
Wt: x.xxxxg
Place filter on balance and
press TARE.
ID
VOLUME REWEIGH
SAMPLE ID:
0
1 2 3 4 5 6 7 8 9 A B
C D E F G H I J K L M N
O P Q R S T U V W X Y Z
Space
Delete
Highlight choice, press ENTER.
Press READY to continue.
➞
➞
➞
➞
XXXXXX
Sample ID:
SAMPLE xx
Power:
xxx %
Temp (xxx): xx.x
Time:
xx:xx
Volume:
xxx
Wt: x.xxxxg
Place filter on balance and press
TARE.
ID
VOLUME REWEIGH
TARING BALANCE . . .