101
Note: Refer to the QUICK TEST section of this
manual for instructions for entering method
parameters for Constant Weight and Set Time.
14.
Press the operation key below NEXT
PAGE to access additional method
parameters.
15.
Press the numbers (1 - 3) and enter the
appropriate method parameters.
16.
Press READY to begin the analysis.
Note: The method test screen appears indicating
the power, temperature, time, and ratio.
17.
If Autotrack is off and a sample identifi-
cation name or number is required, press
the operation key below ID.
18.
Using the operation keys below the arrows,
position the cursor on (select) the first
number or letter of the sample identifi-
cation. Press ENTER. Continue to position
the cursor on each letter or number and
press ENTER until the identification is
complete.
Note: If using an identification number only, use
the numeric keys to enter the number
19.
Press READY to return to the method
screen.
Dilution Moisture/Fat
Fat Parameters
1. POWER: XXX%
2. DRY TIME: XX:XX min:sec
3. BIAS:
±
X.XX%
4. MIN RESULT: X.XX%
5. MAX RESULT: XX.XX%
Press READY to begin test.
MAIN
PREV
NEXT
MENU
PAGE
PAGE
S
E
T
T
I
M
E
Dilution Moisture/Fat
1. MAX TEMP: XXX C
2. MIN WT RANGE: X.XXg
3. MAX WT RANGE: X.XXg
4. WT COMPENSATION: OFF
Press READY to run test.
MAIN
PREV
MENU
PAGE
XXXXXX
Power:
xxx %
Temp (xxx): xx.x
Time:
xx:xx
Ratio:
1.0000
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID
RATIO DILUTION
1 2 3 4 5
SAMPLE ID:
0
1 2 3 4 5 6 7 8 9 A B
C D E F G H I J K L M N
O P Q R S T U V W X Y Z
Space
Delete
Highlight choice, press ENTER.
Press READY to continue.
➞
➞
➞
➞