32
26.
Using the operation keys below the arrows,
position the cursor on (select) the first
number or letter of the sample identifi-
cation. Press ENTER. Continue to position
the cursor on each letter or number and
press ENTER until the identification is
complete.
Note: If using an identification number only, use
the numeric keys to enter the number.
27.
Press READY to return to the Quick Test
screen.
28.
Lift the top cover of the SMART System
5™. Place two glass fiber sample pads on
the balance pan. Close the top cover.
29.
Press TARE.
30.
Wait for the instrument to tare the weight
of the sample pads. The Quick Test screen
will reappear when the tare function is
complete.
Note: The “T” in the lower left corner of the
balance icon indicates that the weight of the
sample pads has been tared.
31.
Lift the instrument cover. Apply the
sample in a thin, even layer to one of the
sample pads. Quickly and gently place the
sample pads back on the balance pan.
Close the top cover.
32.
Press START.
QUICK TEST
Sample ID:
xxxxxx
Power:
xxx%
Temp (xxx): xx.x
Time:
xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID
EDIT
QUICK TEST
Sample ID:
xxxxxx
Power:
xxx%
Temp (xxx):
xx.x
Time:
xx:xx
Wt: x.xxxxg
Place sample and pads on
balance and press START.
ID
EDIT
1 2 3 4 5
TARING BALANCE . . .
T
1 2 3 4 5
SAMPLE ID:
0
1 2 3 4 5 6 7 8 9 A B
C D E F G H I J K L M N
O P Q R S T U V W X Y Z
Space
Delete
Highlight choice, press ENTER.
Press READY to continue.
➞
➞
➞
➞