30
Quick Test
1. POWER: XXX%
2. DELTA WEIGHT: X.X G
3. DELTA TIME: XX secs
4. MAX TIME: XX mins
5. MAX TEMP: XXX C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Input min wt range, press ENTER.
Entry:
02.00
SET
TIME
Quick Test
1. POWER: XXX%
2. DELTA WEIGHT: 0.0 mg
3. DELTA TIME: 0 secs
4. MAX TIME: 10 mins
5. MAX TEMP: 110 C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Input delta time, press ENTER.
Entry:
00
SET
TIME
8.
Press “3” to select and enter delta time.
9.
Using the numeric keys, enter the delta
time
(1 - 60 seconds). Normally a differential
time interval of 10 - 15 seconds is used.
10.
Press ENTER.
Note: Maximum run time is the amount of time
the instrument will operate prior to shutdown if a
constant weight has not been reached. A
maximum run time must be entered or the
program will be invalid. If the instrument shuts
down due to maximum run time, parameters
should be adjusted prior to repeating the
analysis.
11.
Press “4” to select and enter maximum run
time. If using the default value of 10
minutes, omit step 12.
12.
Using the numeric keys, enter the
maximum run time (1 - 60 minutes).
13.
Press ENTER.
14.
Press “5” to select and enter maximum
temperature. If using the default value of
110
°
C, omit step 15.
15.
Using the numeric keys, enter the
maximum temperature (0 - 300
°
C) of the
instrument during the analysis.
16.
Press ENTER.
17.
Press “6” to enter a minimum weight
range. If using the default minimum
weight range of 2.00 g, omit step 18.
18.
Using the numeric keys, enter the
minimum weight range of the sample (0 -
50 g).
19.
Press ENTER.
Quick Test
1. POWER: XXX%
2. DELTA WEIGHT: X.X mg
3. DELTA TIME: XX secs
4. MAX TIME: 10 mins
5. MAX TEMP: 110 C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Input max time, press ENTER.
Entry:
10
SET
TIME
Quick Test
1. POWER: XXX%
2. DELTA WEIGHT: X.X mg
3. DELTA TIME: XX secs
4. MAX TIME: XX mins
5. MAX TEMP: 110 C
6. MIN WT RANGE: 2.00 g
7. MAX WT RANGE: 4.00 g
8. WT COMPENSATION: OFF
Input max temp, press ENTER.
Entry:
110
SET
TIME
➞
➞
➞
➞