106
AVERTISSEMENT
L
’
installation incorrecte du filtre rond
(coussinet) situé sur la partie supérieure
de la chambre d
’
extraction peut exposer
l
’
opérateur à une possible fuite ou vapo-
risataion du solvent. De plus, le systeme
d
’
extraction des matières grasses peut
ètre endommagé résultant du passage
des matières solides dans le système de
distillation.
40.
Press the Run/Stop button on the extractor
to start the extraction cycle. When the
extraction is complete, a beep tone will
sound, and the light in the Run/Stop button
will flash. While the extractor cup is still in
the inverted position, unlatch the lid and
remove the round sample pad on which the
chopped square pads and extracted and
dried sample have been deposited. Latch
the cup lid. The extractor cup will automat-
ically rotate 180 degrees to the upright
position.
Note: To dry additional samples, place the
square and round sample pads on the balance and
press TARE.
41.
Press the operation key below REDRY.
42.
Using the numeric key(s), press the number
of the sample to be redried.
43.
Lift the cover of the SMART System
5
.
Place the round pad with the extracted
sample on the balance pan. Close the
instrument cover.
44.
Press START.
As the analysis begins, the flashing microwave
indicators in the balance icon appear and the
time begins counting down. The sample weight
decreases as the moisture is removed.
Note: Press the operation key below STOP
TEST to end the analysis.
Select Sample ID
1. XXXX XXXX
2. XXXXXX XX
3. XXXXXX
4. XXX
Press item number to select.
XXXXXX
Power:
xxx %
Temp (xxx):
xx.x
Time:
xx:xx
Ratio:
x.xxxx
Wt: x.xxxxg
Place pads on balance and press
TARE or press START to redry.
ID
REDRY
1 2 3 4 5
XXXXX
Sample ID:
xxxxxxxx
Power:
xxx%
Temp (xxx):
xx.x
Redry Time:
xx:xx
Fat: xx.xx%
Wt: x.xxxxg
STOP
TEST