112
20.
Lift the cover of the SMART System
5
.
Place one round and two square glass fiber
sample pads on the balance pan (round pad
on bottom). Close the instrument cover.
21.
Press TARE.
22.
Wait for the instrument to tare the weight
of the pads.
23.
Lift the cover of the SMART System
5
.
Remove the square sample pads from the
balance pan.
24.
Apply the sample in a thin, even layer to
one of the square sample pads and cover
the sample with the other square pad.
25.
With the round pad still on the balance pan,
quickly and gently place the square sample
pads back on the balance pan. Close the
instrument cover.
26.
Press READY.
The SMART System
5
weighs and records the
initial weight of the sample.
XXXXXX
Power:
xxx %
Temp (xxx):
xx.x
Time:
xx:xx
Wt: x.xxxxg
Place one round and two square
pads on balance and press TARE.
ID
REDRY
1 2 3 4 5
TARING BALANCE . . .
Modified Moisture/Fat
Spread sample on two square
pads. Place all pads on
balance and press READY.
ABORT
WEIGHING SAMPLE . . .