3
System Installation
1.
Following the instructions provided in the packing carton, carefully remove
the SMART System
5
from its shipping carton and place it on a vibration-free
workbench or laboratory table in a location that:
a. provides at least 8 in (20 cm) open space on each side and 6 in. (15 cm)
open space in the rear of the instrument for ventilation.
b. is free from vibration of large equipment and/or excessive walk-through
traffic.
c
provides a temperature range of 41
°
F (5
°
C) to 104
°
F (40
°
C) and a
humidity range of 10-85 percent relative humidity.
d. provides adequate bench space for sample handling.
e. allows the instrument to be connected to a dedicated, grounded 120 or
240 VAC outlet. The SMART System
5
™ should be operated on a
stabilized, constant voltage AC power supply. To operate properly, the
voltage must be within
±
10% of the specified level.
2.
Retain all packing materials.
3.
Inspect the instrument for cracks, dents or warping.
4.
Remove the tape from the instrument cover. Inspect the top cover for proper
alignment. The cover must open and close freely with no binding or
restriction of movement.
WARNING
If damage is noted, do not attempt instrument operation.
AVERTISSEMENT
Ne pas mettre en marche si l
’
instrument est endommagé.
5.
If the instrument has been damaged in shipping, contact the freight
carrier to report the damage and to file a damage report. Contact the
CEM Service Department or the nearest subsidiary or distributor (page 4)
to request service information.
6.
Verify that all accessories illustrated and listed in figure 1 are included.
7.
Lift the instrument cover. Install the balance stem into the balance stem
retainer in the opening of the cavity floor, ensuring that the balance stem is
firmly seated. Install the balance pan onto the stem.
Note: If the balance stem requires removal, lift the stem straight up from its
installed position. Moving the balance stem from side to side can cause damage to
the balance assembly.