The instrument reads and records the initial
weight of the sample prior to beginning the
analysis.
As the analysis begins, the flashing microwave
indicators in the balance icon appear and the
time begins counting upward. The sample weight
decreases as the moisture is removed.
During the drying time, the operation key below
SOLIDS or MOISTURE may be pressed to
display % solids or % moisture.
When the weight loss is equal to or less than the
selected parameter, five short beeps will be
heard, and the Data Results screen will appear,
displaying either the % moisture or % solids as
selected.
Note: Press READY to analyze additional
samples using the same method.
Note: Press the operation key below FORM
FEED to advance paper from the internal printer.