6
8.
Using a voltage meter, measure the voltage of the dedicated power outlet to
verify voltage (120 VAC, 60 Hz or 220/240 VAC, 50 Hz).
9.
Facing the back of the instrument, use a small screwdriver to pry the dual
voltage supply from the power module.
10.
Install the voltage slide switch in the proper direction for the measured
voltage (120V, 60Hz or 220/240V, 50Hz).
11.
Based on the voltage selected in step 9, install the proper fuses in the dual
voltage supply – 10 AMP (60 Hz) or 5 AMP (50 Hz).
12.
Install the dual voltage supply with the appropriate fuses and the properly
installed voltage slide switch in the power module.
13.
If applicable, install the exhaust hose.
14.
Plug the power cord into the power cord receptacle in the rear of the
instrument and into a grounded, dedicated electrical outlet. Note: If using
external components such as a printer, computer or balance, install these
components as outlined in this manual.
15.
Position the power switch located in the rear of the instrument above the
power cord receptacle in the “on” position.
CAUTION
Permit the SMART System5™ to warm up (power switch on,
microwaves not being produced) for at least 45 minutes prior to
operation. If the instrument is switched off for an extended
period of time (weekend, holidays, etc.), ensure that warmup
time is permitted prior to use.
Mise en garde
Permettre une période de 45 minutes de réchauffement du
SMART System5 avant l
’
utilisation (interrupteur de courant en
marche, aucune micro-onde n’est émise). Lorsque l
’
instrument
est en mode fermé pour un temps prolongé (fin de semaine,
vacances, etc. . .) il est important que la période de
réchauffement de 45 minutes soit observée.
16.
Refer to the “Maintenance, Troubleshooting And Service” section of this
manual and perform the Standard Solution Test at least five times.
Note: CEM Corporation recommends that the power cord be connected to an AC
power outlet and the power switch remain in the “on” position at all times. After
15 minutes of idle time, the SMART System
5
assumes a “sleep” mode.