![Capintec CAPTUS 2000 Скачать руководство пользователя страница 143](http://html1.mh-extra.com/html/capintec/captus-2000/captus-2000_owners-manual_494180143.webp)
CAPINTEC, INC
CAPTUS
®
2000
Figure 8-44
To proceed with the patient’s plasma sample count, place the plasma sample in the well
counter and select the
OK
button. To cancel the procedure, select the
Cancel
button, and
Figure 8-40 will appear. After the count is completed, Figure 8-39 will appear and require
confirmation of the data.
Once all of the samples counts have been completed, Figure 8-37 will appear again.
PROCEDURE COMPLETE will appear in blue at the bottom of the spectral display area, and
only the
EXIT
button will be active. Choose the
EXIT
button to return to the Main Blood
Volume Test screen.
Test Results
After completion of the measurement process, all of the raw count data and calculated test
results will appear in the appropriately labeled boxes on the patient data screen.
When using the Cr51 test protocol, results are calculated as follows:
(
)
[
]
(
)
C
-
1
L
-
S
A
D
C
H
-
1
P
-
W
Volume
Cell
Red
×
×
×
×
×
=
C
Volume
Cell
Red
Volume
Blood
Whole
=
Volume
Cell
Red
-
Volume
Blood
Whole
Volume
Plasma
=
where W = Net whole blood standard count rate
P = Net plasma standard count rate
H = Hematocrit of standard
C = Hematocrit of patient
D = Administered to standard ratio
A = Administered Volume
S = Net patient whole blood
sample count rate
L = Net patient plasma sample count rate
July 00
LAB TESTS
8-35
Содержание CAPTUS 2000
Страница 2: ......
Страница 8: ......
Страница 26: ......
Страница 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Страница 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Страница 52: ......
Страница 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Страница 66: ......
Страница 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Страница 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Страница 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Страница 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Страница 158: ......
Страница 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Страница 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Страница 174: ......
Страница 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Страница 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Страница 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Страница 206: ......