73
LOG MESSAGES
Message
Description
Add Air
Additional air cycle was run
Add Air On
Air cycle will be added at the end of the current cycle
Add Air Off
Air cycle will not be added at the end of the current cycle
Air
Indicates when the air portion of a cycle occurred
Auto Dis
Auto water line disinfectant cycle was run
Disinfect
Indicates when the disinfectant portion of a cycle occurred
HLD Fail
Disinfectant efficacy has failed MRC test
HLD Pass
Disinfectant efficacy has passed MRC test
Leak Disabled
Leak tester was disabled
Leak Fail
Leak test failed
Leak Test
Indicates when the leak test portion of a cycle occurred
Leak Sen Err
The leak tester measured pressure at the start of a cycle
Low Alcohol
Alcohol level is low
Machine ID
The last 4 digits of the DSD EDGE™ Reprocessor serial number
No Air Flow
No air flow was detected
No Fluid Flo
No fluid flow was detected
Power on
The device lost power during a cycle
Resume
Indicates a cycle resumed