25
REPROCESSING CYCLE
•
If monitored sensors are not satisfied during startup, the cycle is halted until the
issue is resolved or cycle is cancelled.
•
Automated Leak Tester (Optional).
If enabled, a 40 second leak test is
performed before starting the Flush phase.
•
Fresh water is flushed through scope channels and basin.
•
Detergent (Optional).
If programmed, detergent is injected into the
channels prior to the fresh water flush.
•
Incoming water temperature is monitored to ensure water temperature is reaching
proper range for disinfection.
•
Concludes by purging scope channels with air and emptying the basin of water.
•
An additional one or two soak segments can be added to the Flush phase.
•
No detergent is injected during additional soak segments.
•
Endoscope channels are flushed with disinfectant.
•
Basin is filled with disinfectant solution and endoscope is soaked for the 5 minute
disinfectant soak time.
•
Concludes by draining disinfectant from the basin.
•
Basin is partially filled with fresh water as scope channels are flushed with fresh
water. Partially filled basin is then drained.
•
Basin is completely filled with fresh water as scope channels are flushed with
fresh water. Basin is then drained while the channels are purged with fresh water.
•
Concludes by purging scope channels with air.
STARTUP
FLUSH
WASH
(Optional)
DISINFECT
RINSE 1