CHECK SYSTEM TIME
1. On the control panel, check the time for accuracy.
2. If the system time is incorrect, reset the time (for complete instructions, see
Additional Operation Functions
).
CHECK PRE-FILTERS
1. With water flowing through the filters, inspect the
pressure gauges for pressure reading.
2. Replace any filters if the dynamic pressure drop
across any filter is greater than 5 psi (for filter
replacement instructions, see
Periodic Maintenance
).
Figure 4.1
- Inspect Pressure Gauges for
Pressure Drop
Note:
The personnel responsible for these tasks varies by location. Verify your location’s
policies and procedures before performing any of the tasks in this chapter.
Note:
When performing maintenance activities on the DSD EDGE™ Reprocessor,
ensure there is adequate ambient illumination to prevent against hazards during maintenance.