40
|
MANUALLY INTERRUPT THE CYCLE
If necessary, a cycle can be manually interrupted at any time.
1. On the control panel, verify the correct station is selected. If necessary, press
STATION SELECT
to choose
the station (A or B).
2. Press
STOP
.
COMPLETE THE DISINFECTION PROCESS
Note:
Intended to be used with RAPICIDE™ PA High-Level Disinfectant and RAPICIDE™ PA
High-Level Disinfectant Test Strips.
Caution!
The leak test adapter must be disconnected and removed from the basin when
not in use to avoid potential fluid ingression.
When the disinfection process is complete, the LCD screen (and optional
alarm) prompts the operator to verify the disinfectant minimum recommended
concentration has been met by following the RAPICIDE™ PA High-Level
Disinfectant Test Strips Instructions for Use.
1. Dip a test strip into the disinfectant sample port located near the front-left
corner of the basin.
2. Compare the results to the chart provided on the test strip bottle.
3. Do one of the following:
a. If the test strip color indicates the disinfectant minimum recommended
concentration was met for the cycle, press
HLD PASS
on the
control panel.
After ten seconds, the process indicator light illuminates and “Completed”
displays on the screen.
b. If the test strip color indicates the disinfectant minimum recommended
concentration was not met during the cycle, press
HLD FAIL
on the
control panel.
Figure 5.3
- Dip Test Strip and
Compare Results to Chart
If the minimum recommended concentration was not
met, the endoscope is not patient safe. Stop here and
re-run the disinfection cycle for the endoscope.